PET systems


Scanner type

Studied FOM

Agreement

References

ECAT EXACT HR+, CPS

Spatial resolution
Sensitivity
Count rates


Scatter fraction

about 3 %
< 7 %
good at activity
concentrations < 20 kBq/ml
about 3 %

Jan et al 2005

ECAT HRRT, Siemens

Spatial resolution
Scatter fraction
Scattered coinc profiles
Count rates

excellent (<0.2 mm )
< 1 %
very good (visual)
good (about 10%)

Bataille et al 2004

Hi-Rez, Siemens

Scatter fraction
Count rates

NEC curves

about 1 %
good at activity
concentrations < 40 Bq/ml
good at activity
concentrations < 40 Bq/ml

Michel et al 2006

Allegro, Philips

Count rate
Scatter fraction

< 8 %
8 %

Lamare et al 2006

GE Advance, GEMS

Energy spectra
Scatter fraction

not reported
< 1 %

Schmidtlein et al 2006

MicroPET P4, Concorde

Spatial resolution
Sensitivity
Miniature Derenzo phantom

about 7 %
< 4 %
visual assessment

Jan et al 2003

MicroPET Focus 220, Siemens

Spatial resolution
Sensitivity
Count rates for mouse phantom

about 5 %
about 3 %
prompt coinc: < 5.5 %
delayed coinc: < 13 %

Jan et al 2005

Mosaic, Philips

Scatter fraction
Count rates

about 5 %
4-15 %

Merheb et al 2005